The Evaluating the sensitivity of Hexagon OBTI test and Kastel-Meyer test by detection of human blood

Authors

  • Yousif M. Jebur Najaf Governorate Police Command, Criminal Evidence Department

DOI:

https://doi.org/10.36320/ajb/v16.i1.13363

Keywords:

Hexagon OBTI. Kastel-Meyer. Detection of human blood.

Abstract

The Hexagon OBTI test is a commercial kit made for the purpose of forensically identifying human blood. The test process is straightforward and does not need for a lab environment. To reduce between-sampling variation, all samples (blood) were taken from the researcher, a healthy male who was 33 at the time of collection. Blood thinners were created using liquid blood supplies and deionized water in the following dilutions: 1:10, 1:100, 1:1000, 1:100,000, 1:1,000,000, 1:10000000, and 1:100000000. The samples were analyzed by Hexagon OBTI and Kastel-Meyer, and the findings of both tests were positive up to a dilution of 1:1000 in the case of the former and 1:1000000 in the case of the latter. The result was a more sensitive OBTI test than Kastel-Meyer , the OBTI test is therefore suitable for identification of human blood at crime scenes 

 

Downloads

Download data is not yet available.

References

Jackson, A.R.W. & Jackson, J.M. (2010). Forensic Science, 3rd Edition. Essex, Pearson Education Ltd.

Frumkin D, Wasserstrom A, Davidson A, et al.Authentication of forensic DNA samples. Forensic SciInt Genet 2010; 4(2): 95–103. DOI: https://doi.org/10.1016/j.fsigen.2009.06.009

Muro CK, Doty KC, de Souza Fernandes L, et al.Forensic body fluid identification and differentiation byRaman spectroscopy. Forensic Chem 2016; 1: 31–38. DOI: https://doi.org/10.1016/j.forc.2016.06.003

Hochmeister, M.; Sparkes, R.; Rudin, O.; Gehrig, C.; Schmidt, L., Cordier, A.; Dirnhofer, R. Evaluation of an Immunochromatographic 1-step Test for the Forensic Identification of Human Blood. Presented at the 50th annual meeting of the American Academy of Forensic Sciences, San Francisco, CA, 1998. DOI: https://doi.org/10.1520/JFS14516J

Hermon, D., Shpitzen, M., Oz, C., Glattstein, B., Azoury, M., & Gafny, R. (2003). The Use of the Hexagon OBTI Test for Detection of Human Blood at Crime Scenes and on Items of Evidence Part 1: Validation Studies and Implementation. Journal of Forensic Identification, 53(5), 566-575.

Cox, M. A. Study of the Sensitivity and Specificity of Four Presumptive Tests for Blood. J. For. Sci. 1991, 36 (5), 1503- 1511. DOI: https://doi.org/10.1520/JFS13170J

https://www.bluestar-forensic.com/medias/ documentationfile/bluestar_obti_verification_report.pdf

K. Virkler, I.K. Lednev, Analysis of body fluids for forensic purposes: from laboratory testing to non-destructive rapid confirmatory identification at a crime scene, Forensic Sci. Int. 188 (1–3) (2009) 1–17. DOI: https://doi.org/10.1016/j.forsciint.2009.02.013

S.S. Tobe, N. Watson, N.N. Dae´id, Evaluation of six presumptive tests for blood, their specificity, sensitivity, and effect on high molecular-weight DNA, J. Forensic Sci. 52 (1) (2007) 102–109. DOI: https://doi.org/10.1111/j.1556-4029.2006.00324.x

M. Grodsky, K. Wright, P.L. Kirk, Simplified preliminary blood testing. An improved technique and a comparative study of methods, J. Crim. Law Criminol. Police Sci. 42 (1) (1951) 95–104. DOI: https://doi.org/10.2307/1140307

Idris, B., & Goodwin, W. H. (2022). Evaluating the sensitivity of presumptive and confirmatory tests for body fluids. Forensic Science International: Genetics Supplement Series, 8, 276-278. DOI: https://doi.org/10.1016/j.fsigss.2022.10.061